1. Sensis, material # 10764561; UDI: 04056869010137 2. Sensis Vibe Hemo, material # 11007641; UDI: 04056869010199 3. Sensis Vibe Combo, material # 11007642; UDI: 04056869010205 The Sensis recording system is a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac, interventional Radiology, and surgical studies.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    1. Sensis- 10764561; UDI: 04056869010137; Serial Numbers: 103493, 103432, 103466 2. Sensis Vibe Hemo, material # 11007641; UDI: 04056869010199; Serial Numbers: 103137, 103144, 103453, 103108, 103111, 103047, 103048, 103171, 103182, 103193, 103301, 103170, 103223,, 103225, 103247, 103313, 103315, 103196, 103198, 103166, 103173, 103177, 103184, 103186, 103199, 102326, 103139, 103161, 103167, 103168, 103172, 103178, 100263, 103424, 103319, 103320, 103321, 103323, 103325, 103328, 103330, 103333, 103334, 103335, 103344, 100216, 103337, 100416, 102008, 103413, 101083, 100208, 100209, 102138, 102145, 102147, 103160, 103314, 102199, 103079, 102221, 103123, 103124, 103131, 100163, 100306, 100309, 103300, 101099, 101100, 101101, 103265, 100422, 103019, 103331, 103362, 103393, 103402, 103419, 103252, 103261, 103267, 103269, 103270, 103276, 103277, 103280, 103281, 103022, 103381, 103194 103145, 103148, 103152, 100345, 102183, 103345, 103368, 103415, 103416, 103245, 103249, 103250, 103264, 103080, 103082, 103086, 103088, 103089, 103090, 100231, 103431, 103306, 103341, 103114, 100070, 100071, 100074, 102155, 102157, 103040, 103042, 103039, 102174, 102182, 103130, 103200, 103234, 103116, 103117, 103119, 103288, 103008, 103216, 103465, 103189, 103208, 102010, 102295, 102296, 102302, 100006, 100005, 103352, 103283, 103308, 103310, 103115, 103244, 103246, 103014, 103242, 102143, 103359, 103274, 103437 3. Sensis Vibe Combo, material # 11007642; UDI: 04056869010205; Serial Numbers: 123021, 123004, 123047, 123053, 122011 ***Updated 1/18/22*** SENSIS Vibe Hemo, Serial #103185 ***Updated 2/24/22*** SENSIS Vibe Hemo Serial #103290 103510 103451 103461 103444 103450 103448 103445 103442 103449 103439 103457 SENSIS Vibe Combo Serial #123055 123056 123058 The following serial numbers have been removed from the scope of this recall: SENSIS Vibe Hemo Serial #103301 103424 ***Updated 8/5/22*** The following serial numbers have been added to the recall: SENSIS Vibe Combo Serial numbers 103307, 103312
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Siemens Medical Solutions USA, Inc
Reason for Recall:
Software error which affects Sensis Vibe Hemo, Sensis and Sensis Vibe Combo systems with software version VD12A, which could cause possible hazard to patients, operators, or other persons and equipment. Under certain sporadic circumstances, the CO (Cardiac Output) measurement using the Thermodilution method will temporarily no longer be possible.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

1. Sensis, material # 10764561; UDI: 04056869010137 2. Sensis Vibe Hemo, material # 11007641; UDI: 04056869010199 3. Sensis Vibe Combo, material # 11007642; UDI: 04056869010205 The Sensis recording system is a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac, interventional Radiology, and surgical studies.

Product Codes/Lot Numbers:

1. Sensis- 10764561; UDI: 04056869010137; Serial Numbers: 103493, 103432, 103466 2. Sensis Vibe Hemo, material # 11007641; UDI: 04056869010199; Serial Numbers: 103137, 103144, 103453, 103108, 103111, 103047, 103048, 103171, 103182, 103193, 103301, 103170, 103223,, 103225, 103247, 103313, 103315, 103196, 103198, 103166, 103173, 103177, 103184, 103186, 103199, 102326, 103139, 103161, 103167, 103168, 103172, 103178, 100263, 103424, 103319, 103320, 103321, 103323, 103325, 103328, 103330, 103333, 103334, 103335, 103344, 100216, 103337, 100416, 102008, 103413, 101083, 100208, 100209, 102138, 102145, 102147, 103160, 103314, 102199, 103079, 102221, 103123, 103124, 103131, 100163, 100306, 100309, 103300, 101099, 101100, 101101, 103265, 100422, 103019, 103331, 103362, 103393, 103402, 103419, 103252, 103261, 103267, 103269, 103270, 103276, 103277, 103280, 103281, 103022, 103381, 103194 103145, 103148, 103152, 100345, 102183, 103345, 103368, 103415, 103416, 103245, 103249, 103250, 103264, 103080, 103082, 103086, 103088, 103089, 103090, 100231, 103431, 103306, 103341, 103114, 100070, 100071, 100074, 102155, 102157, 103040, 103042, 103039, 102174, 102182, 103130, 103200, 103234, 103116, 103117, 103119, 103288, 103008, 103216, 103465, 103189, 103208, 102010, 102295, 102296, 102302, 100006, 100005, 103352, 103283, 103308, 103310, 103115, 103244, 103246, 103014, 103242, 102143, 103359, 103274, 103437 3. Sensis Vibe Combo, material # 11007642; UDI: 04056869010205; Serial Numbers: 123021, 123004, 123047, 123053, 122011 ***Updated 1/18/22*** SENSIS Vibe Hemo, Serial #103185 ***Updated 2/24/22*** SENSIS Vibe Hemo Serial #103290 103510 103451 103461 103444 103450 103448 103445 103442 103449 103439 103457 SENSIS Vibe Combo Serial #123055 123056 123058 The following serial numbers have been removed from the scope of this recall: SENSIS Vibe Hemo Serial #103301 103424 ***Updated 8/5/22*** The following serial numbers have been added to the recall: SENSIS Vibe Combo Serial numbers 103307, 103312

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0499-2022

Related Recalls

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

Mar 9, 2026 Other Medical Devices View Details →
Class I - Dangerous

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

Mar 9, 2026 Other Medical Devices View Details →

ARTIS Pheno VE30A and VE40A, Model 10849000

Siemens Medical Solutions USA

Class I - Dangerous

During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.

Mar 12, 2026 Other Medical Devices View Details →