RefleXion Medical Radiotherapy System - System Label: "*** refleXion REF RXM1000 ***" Physics Guide/Treatment Delivery User Manual/Treatment Planning User Manual: "RefleXion X1 Model: RXM1000"
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model: RXM1000 Serial Numbers/UDI Codes: X11001 / (01)00860003983805(11)210125(21)X11001 X11002 / (01)00860003983805(11)210324(21)X11002 X11003 / (01)00860003983805(11)210524(21)X11003
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Reflexion Medical, Inc.
- Reason for Recall:
- Due to dose discrepancy when delivering a plan to a patient in a Non-HFS (Head First Supine) orientation specifically in Feet First Supine (FFS).
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
RefleXion Medical Radiotherapy System - System Label: "*** refleXion REF RXM1000 ***" Physics Guide/Treatment Delivery User Manual/Treatment Planning User Manual: "RefleXion X1 Model: RXM1000"
Product Codes/Lot Numbers:
Model: RXM1000 Serial Numbers/UDI Codes: X11001 / (01)00860003983805(11)210125(21)X11001 X11002 / (01)00860003983805(11)210324(21)X11002 X11003 / (01)00860003983805(11)210524(21)X11003
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0470-2022
Related Recalls
RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00
Reflexion Medical
Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.
Due to incorrect software configuration that potentially allows more than two (2) fractions within a 12-hour period (fractionLimits) and could potentially lead to adverse events (toxicity).
A potential dose error exists for patients treated with an out of session SCINTIX partial fraction.