SCFE Driver (SCF-MLD273 & SCF-MLD265) instrument used for the insertion of the Free Gliding SCFE Screw System. This instrument is included in the Free Gliding SCFE Screw System instrument tray. This instrument is marked with the catalogue number and lot number. The Free-Gliding SCFE Screw System is a temporary implant for stabilization of pediatric femoral neck fractures and slipped capital femoral epiphysis (SCFE) in pediatric patients. Orthopedic.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Part Number: SCF-MLD265; LOT#: 130703-01 Part Number: SCF-MLD273; LOT#: 130703-02
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Pega Medical Inc.
- Reason for Recall:
- SCFE Driver's threaded shaft broke close to the knob when a manual force was applied during the fixation of the screw.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
SCFE Driver (SCF-MLD273 & SCF-MLD265) instrument used for the insertion of the Free Gliding SCFE Screw System. This instrument is included in the Free Gliding SCFE Screw System instrument tray. This instrument is marked with the catalogue number and lot number. The Free-Gliding SCFE Screw System is a temporary implant for stabilization of pediatric femoral neck fractures and slipped capital femoral epiphysis (SCFE) in pediatric patients. Orthopedic.
Product Codes/Lot Numbers:
Part Number: SCF-MLD265; LOT#: 130703-01 Part Number: SCF-MLD273; LOT#: 130703-02
Distribution:
Distributed in: FL, TN
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0324-2016
Related Recalls
The lot number on the pediatric orthopedic implant driver instrument handle does not match the lot number on the label.
There is a potential for the screws to have become loosened during ultrasonic cleaning of the knobs and the screws may come out.
SLIM Handle - catalogue #: SLM-HND100 a component of Simple Locking Intramedullary (SLIM) System
Pega Medical
Handle could jam with the Driver due to the detachment of an internal Connector Ring.