9" (23 cm) Appx 0.63 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, NanoClave, 3 Clamps, 0.2 Micron Filter, Rotating Luer; 1 unit per pouch, 50 multivac pouches per case UDI:(01)10887709052953(17)241201(30)50(10)4536611;(01)10887709052953(17)250101(30)50(10)4557492 The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Numbers:4536611, 4557492.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- ICU Medical, Inc.
- Reason for Recall:
- Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
9" (23 cm) Appx 0.63 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, NanoClave, 3 Clamps, 0.2 Micron Filter, Rotating Luer; 1 unit per pouch, 50 multivac pouches per case UDI:(01)10887709052953(17)241201(30)50(10)4536611;(01)10887709052953(17)250101(30)50(10)4557492 The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).
Product Codes/Lot Numbers:
Lot Numbers:4536611, 4557492.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0189-2021
Related Recalls
Plum Duo Precision IV Pump, 40002-0403
ICU Medical
Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.
ICU Medical Plum Duo Infusion Pump, List Numbers: 400020401 (US), 400021001 (Philippines)
ICU Medical
Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.
Tego Connector: REF: 011-D1000, 01C-D1000, 055-D1000, D1000; Conector Tego: REF: LAT-D1000
ICU Medical
Needle free access device intended as accessory to vascular access used in hemodialysis or accessory to intravascular administration sets has a silicone seal which may bulge at the top, separate from the body, tear; may result in fluid leaks, occluded fluid flow, inability to inject/withdraw blood product, delayed/interrupted therapy, exposure to biological contaminants, air infused into the body