Atellica IM HER-2/neu (H2n) assay, 50 Test Kit, Catalog No. SMN 10995591, UDI Codes: (01)00630414599151(10)98034139(17)20210130 and (01)00630414599151(10)32478139(17)20210130 Atellica IM HER-2/neu (H2n) calibrator, 2 Pack, Catalog No. SMN 10995592, UDI Code (01)00630414599168(10)22695A49(17)20201211 ADVIA Centaur HER-2/neu (H2n) assay, 50 Test Kit, Catalog No. SMN 10308994, UDI codes : (01)00630414471235(10)43157138(17)20210130 (01)00630414471235(10)81653138(17)20210130 (01)00630414471235(10)99728138(17)20210130 (01)00630414471235(10)24043138(17)20210130 (01)00630414471235(10)32477138(17)20210130 Advia Centaur Systems HER-2/neu (H2n) calibrator, 2 Pack, Catalog No. SMN 10308993, UDI Codes: (01)00630414471228(10)64661A48(17)20201112 (01)00630414471228(10)81211A48(17)20201112 (01)00630414471228(10)91743A48(17)20201112 (01)00630414471228(10)22124A49(17)20201211 Product Usage: This product is intended to be used for in vitro diagnostic use in calibrating the ADVIA Centaur systems HER-2/neu assay.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Atellica IM HER-2/neu Assay 50 Test Kit: Siemens Material Number (SMN) 10995591; EXP Date 2021-01-30; Lot # 98034139 w/ UDI (01)00630414599151(10)98034139(17)20210130; Lot # 32478139 w/ UDI (01)00630414599151(10)32478139(17)20210130 Atellica IM HER-2/neu Calibrator (2 pack): SMN 10995592; EXP Date 2020-12-11; Lot # 22695A49; UDI (01)00630414599168(10)22695A49(17)20201211 ADVIA Centaur HER-2 Assay 50 Test Kit: SMN 10308994; EXP Date 2021/01/30; Lot # 43157138 w/ UDI (01)00630414471235(10)43157138(17)20210130 ; Lot# 81653138 w/ UDI (01)00630414471235(10)81653138(17)20210130 ; Lot# 99728138 w/ UDI (01)00630414471235(10)99728138(17)20210130 ; Lot # 24043138 w/ UDI (01)00630414471235(10)24043138(17)20210130 ; Lot #32477138 w/ UDI (01)00630414471235(10)32477138(17)20210130 ; ADVIA Centaur HER-2/neu Calibrator (2 pack): SMN 10308993; EXP Date 2020-11-12 Lot # 64661A48 w/ UDI (01)0063041447(10)64661A48(17)20201112 Lot # 81211A48 w/ UDI (01)0063041447(10)81211A48(17)20201112 Lot# 91743A48 w/ UDI (01)0063041447(10)91743A48(17)20201112 EXP Date 2020-12-11; Lot # 22124A49 w/ UDI (01)0063041447(10)22124a49(17)20201112
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Siemens Healthcare Diagnostics, Inc
Reason for Recall:
Siemens confirmed that the Upper Limit of Normal (ULN) as claimed in the Instructions for Use (IFU) was no longer achieved. This can potentially result in false positive or false negative results.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Atellica IM HER-2/neu (H2n) assay, 50 Test Kit, Catalog No. SMN 10995591, UDI Codes: (01)00630414599151(10)98034139(17)20210130 and (01)00630414599151(10)32478139(17)20210130 Atellica IM HER-2/neu (H2n) calibrator, 2 Pack, Catalog No. SMN 10995592, UDI Code (01)00630414599168(10)22695A49(17)20201211 ADVIA Centaur HER-2/neu (H2n) assay, 50 Test Kit, Catalog No. SMN 10308994, UDI codes : (01)00630414471235(10)43157138(17)20210130 (01)00630414471235(10)81653138(17)20210130 (01)00630414471235(10)99728138(17)20210130 (01)00630414471235(10)24043138(17)20210130 (01)00630414471235(10)32477138(17)20210130 Advia Centaur Systems HER-2/neu (H2n) calibrator, 2 Pack, Catalog No. SMN 10308993, UDI Codes: (01)00630414471228(10)64661A48(17)20201112 (01)00630414471228(10)81211A48(17)20201112 (01)00630414471228(10)91743A48(17)20201112 (01)00630414471228(10)22124A49(17)20201211 Product Usage: This product is intended to be used for in vitro diagnostic use in calibrating the ADVIA Centaur systems HER-2/neu assay.

Product Codes/Lot Numbers:

Atellica IM HER-2/neu Assay 50 Test Kit: Siemens Material Number (SMN) 10995591; EXP Date 2021-01-30; Lot # 98034139 w/ UDI (01)00630414599151(10)98034139(17)20210130; Lot # 32478139 w/ UDI (01)00630414599151(10)32478139(17)20210130 Atellica IM HER-2/neu Calibrator (2 pack): SMN 10995592; EXP Date 2020-12-11; Lot # 22695A49; UDI (01)00630414599168(10)22695A49(17)20201211 ADVIA Centaur HER-2 Assay 50 Test Kit: SMN 10308994; EXP Date 2021/01/30; Lot # 43157138 w/ UDI (01)00630414471235(10)43157138(17)20210130 ; Lot# 81653138 w/ UDI (01)00630414471235(10)81653138(17)20210130 ; Lot# 99728138 w/ UDI (01)00630414471235(10)99728138(17)20210130 ; Lot # 24043138 w/ UDI (01)00630414471235(10)24043138(17)20210130 ; Lot #32477138 w/ UDI (01)00630414471235(10)32477138(17)20210130 ; ADVIA Centaur HER-2/neu Calibrator (2 pack): SMN 10308993; EXP Date 2020-11-12 Lot # 64661A48 w/ UDI (01)0063041447(10)64661A48(17)20201112 Lot # 81211A48 w/ UDI (01)0063041447(10)81211A48(17)20201112 Lot# 91743A48 w/ UDI (01)0063041447(10)91743A48(17)20201112 EXP Date 2020-12-11; Lot # 22124A49 w/ UDI (01)0063041447(10)22124a49(17)20201112

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0115-2021

Related Recalls

Class I - Dangerous

Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results

Mar 18, 2026 Other Medical Devices Nationwide View Details →

A potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP on separate analyzers are not impacted by the recall. The frequency and magnitude of the negative bias and imprecision increase over time and are not consistent with the assay Instructions For Use (IFU). When this issue is present, Quality Control (QC) results demonstrate a negative trend and increased imprecision. The clinical impact of a negative (A1c_E) bias may lead to a delay in diagnosis for diabetes or lead to less-than-optimal treatment for existing diabetes.

Feb 23, 2026 Other Medical Devices Nationwide View Details →