Coconut Bun, NET WT. 4 OZ. Product is packaged in an uncoded, plastic bag with green font. MADE BY GOLDEN KING BAKERY, INC. 90 Bowery New York, NY 10013 212-680-0138.
Class I - DangerousWhat Should You Do?
- Check if you have this product: none
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Golden King Bakery
- Reason for Recall:
- The product was found to contain Undeclared Egg based on samplings and analyses by New York State Department of Agriculture & Markets
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Coconut Bun, NET WT. 4 OZ. Product is packaged in an uncoded, plastic bag with green font. MADE BY GOLDEN KING BAKERY, INC. 90 Bowery New York, NY 10013 212-680-0138.
Product Codes/Lot Numbers:
none
Distribution:
Distributed in: NY, CT, PA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: F-0924-2015