Penner Pacific Bathing Spa, Model Numbers 560010-2L, 560010-XL, 560010-1L, 560010-XL
Penner Patient Care
The device does not bear a unique device identifier.
Medical equipment, implants, and diagnostic devices
Penner Patient Care
The device does not bear a unique device identifier.
Product not cleared by the FDA.
Becton Dickinson & Company
During the bulk packaging process, some Luer Slip syringes were incorrectly included in a batch designated for Luer Lok syringes.
Olympus Corporation of the Americas
Potential for rubber fragment detachment during use.
Differential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift.
Olympus Corporation of the Americas
Potential for rubber fragment detachment during use.
Differential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift.
Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.
The potential for the needle tip to be dull/blunt, difficult to advance, and/or break.
Differential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift.
Staar Surgical AG
Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.
There is the potential for the length of the trocar shaft to be too long.
A potential weakness has been identified in the probe body, so that, in a specific production batch, liquid may leak from the terminal part of the casing near the cable.
Software defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assays. Cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semi-quantitative and quantitative clinical chemistry and immunochemistry assays, as well as ion-selective measurements. When an erroneous calibration is active, the instrument fails to calculate new values and instead repeats the last successfully calculated result from any Spline-type assay for all subsequent measurements, leading to identical and erroneous patient and QC results. The defect could lead to patients receiving erroneous lab results, which could result in a remote risk of serious adverse health consequences. The following assays use spline type calibration and are affected by this issue: Cystatin C Gen.2, Ferritin Gen.4, Lipoprotein (a) Gen. 2, Lipoprotein (a) molarity, Vancomycin Gen.3, Kappa Free Light Chains Partner Channel, Lambda Free Light Chains Partner Channel, and fCAL turbo Partner Channel.
Under certain workflows, patient information shown in the viewer may not match the images displayed on Centricity Universal Viewer Zero Footprint Client (ZFP) with versions ZFP v6.0: SP11.6, SP11.6.0.1 and SP11.6.1 on systems with Patient Timeline installed,
The firm has identified that the battery runtime and cycle specifications contained in the devices Instructions for Use (IFU) need to be updated.
Siemens Healthcare Diagnostics
Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.
The firm has identified that the battery runtime and cycle specifications contained in the devices Instructions for Use (IFU) need to be updated.
The potential for the needle tip to be dull/blunt, difficult to advance, and/or break.
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.