The torque limiting adapter may not stay locked inside of the ratcheting handle once fully engaged.
Surgical Instruments
🏥 Medical Devices • 7,015 recalls
Three lots tied by two trainees were released without performing the required pull test. If tied with inadequate strength, the proximal ligature could slip toward the tip of the catheter and block the inflation hole, leading to the risk of the tip breaking off, and potentially becoming lost in the blood vessel.
Two electronic diode components (D1 & D4) placed in the reverse orientation on the controller printed circuit board assembly may result in premature wear of the power regulator, leading to the regulators failure which could cause a full power loss. If the regulator failure were to occur during a patient procedure, a procedural delay may result
Due to a defect in the outer pouch sterile seal
Procedure Instruments with erroneous unit carton labels.
GentleWave System Posterior CleanFlow Procedure Instrument with Matrices (Single) REF GW-PST-PI01
Sonendo
Procedure Instruments with erroneous unit carton labels.
Scalpel and StatLock included with HydroPICC Catheter Kit - PICC-142, Peripherally Inserted Central Catheter
Access Vascular
Scalpel and StatLock components packaged with specific lots of the HydroPICC catheter have a shorter expiration date than what is listed on the outer kit package.
Due to manufacturing instruments which included additional features and adequate consideration was not given to the need for a new Regulatory filing.
the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.
THERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S. Used for catheter-based cardiac electrophysiological mapping (stimulation and recording) and, in conjunction with an RF generator, for cardiac ablation.
Johnson & Johnson Surgical Vision
Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation causing catheter tip temperature increase, limiting the ability to deliver ablation, or 3) particles flushed through the catheter irrigation may have thrombogenic/toxic effect.
the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.
Due to an increase of confirmed latex deterioration complaints for arterial embolectomy catheters packaged in pouches.
BAUSCH + LOMB BASIC OPHTHALMIC PACK, REF LYN023BSBLB. Convenience kit used in surgical procedures.
MEDLINE INDUSTRIES, LP - Northfield
Medline received complaints on some of the breather pouches with in the sterile surgical kits show openings in the seal.
BAUSCH + LOMB CATARACT TRAY, REF LYN023FACTB. Convenience kit used in surgical procedures.
MEDLINE INDUSTRIES, LP - Northfield
Medline received complaints on some of the breather pouches with in the sterile surgical kits show openings in the seal.
MEDLINE LITHOTOMY, REF DYNJ50099C. Convenience kit used in medical/surgical procedures.
MEDLINE INDUSTRIES, LP - Northfield
Medline received complaints on some of the breather pouches with in the sterile surgical kits show openings in the seal.
BAUSCH + LOMB PHACO PACK, REF LYN011PHNHD. Convenience kit used in surgical procedures.
MEDLINE INDUSTRIES, LP - Northfield
Medline received complaints on some of the breather pouches with in the sterile surgical kits show openings in the seal.
BAUSCH + LOMB EYE PACK, REF LYN003EYSPA. Convenience kit used in surgical procedures.
MEDLINE INDUSTRIES, LP - Northfield
Medline received complaints on some of the breather pouches with in the sterile surgical kits show openings in the seal.
BAUSCH + LOMB CATARACT PACK, REF VAL001CPMEA. Convenience kit used in surgical procedures.
MEDLINE INDUSTRIES, LP - Northfield
Medline received complaints on some of the breather pouches with in the sterile surgical kits show openings in the seal.
BD Nexiva Closed IV Catheter System - Single Port REF 383516 20 GA X1.0o in (1.1 x 25 mm) 61 mL/min (3660 mL/hr)
Becton Dickinson Infusion Therapy Systems
During needle withdrawal, the tip shield may prematurely separate from the catheter adapter and preventing the needle tip from being in the safely shielded (leaving the needle tip exposed).
MEDLINE LINQ TRAY, REF DYNJ65880A. Convenience kit used in medical/surgical procedures.
MEDLINE INDUSTRIES, LP - Northfield
Medline received complaints on some of the breather pouches with in the sterile surgical kits show openings in the seal.