Stryker Neurovascular is recalling their DAC- Distal Access Catheter -038 136cm, a Distal Access Catheter by removal. The reason for the recall is: the DAC Lot #0000486382 was released with out-of-specification endotoxin results.
Surgical Instruments
🏥 Medical Devices • 7,015 recalls
POWERSEAL 5 mm, 44 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascular, thoracic, and gynecologic. Model/Catalog Number: PS-0544CJDA
Olympus Corporation of the Americas
Upon initial activation of the seal function, the POWERSEAL non-conformance results in an immediate Incomplete Seal Cycle tone with the accompanying message on the generator screen, and no energy would be delivered to the device resulting in delay in surgery.
One lot of product was exposed to more than the approved number of Ethylene Oxide (EO) sterilization cycles. This may decrease the tensile strength of the sutures over time, which could result in harms such as wound dehiscence, hemorrhage, blood loss, bleeding, tissue breakdown, peritonitis, unspecified infection, vision loss (when used in ophthalmic application), and/or prolonged surgery.
POWERSEAL 5 mm, 23 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascular, thoracic, and gynecologic. Model/Catalog Number: PS-0523CJDA
Olympus Corporation of the Americas
Upon initial activation of the seal function, the POWERSEAL non-conformance results in an immediate Incomplete Seal Cycle tone with the accompanying message on the generator screen, and no energy would be delivered to the device resulting in delay in surgery.
POWERSEAL 5 mm, 37 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascular, thoracic, and gynecologic. Model/Catalog Number: PS-0537CJDA
Olympus Corporation of the Americas
Upon initial activation of the seal function, the POWERSEAL non-conformance results in an immediate Incomplete Seal Cycle tone with the accompanying message on the generator screen, and no energy would be delivered to the device resulting in delay in surgery.
Route 92 Medical, Tenzing 7 Delivery Catheter, REF 1005, Rx Only, Sterile EO. Used for introduction of interventional devices into the neurovasculature.
ROUTE 92 MEDICAL
Their is the potential for distal tip separation at the proximal marker band of the delivery catheters
Products may not be heparin coated as indicated on the label-leads to a potential loss of benefit due to the absence of the heparin coating
3M Durapore Surgical Tape, Catalog Number 1538-118
3M Company - Health Care Business
Select lots of the 3M Durapore Surgical Tape, Catalog Number 1538-118, were incorrectly labeled with a shelf life of 5 years, rather than 3 years.
Due to complaints of particulates, residue, occlusions, and bent tips for sterile cannulas.
Products may not be heparin coated as indicated on the label-leads to a potential loss of benefit due to the absence of the heparin coating
Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.
The Surgical Tools and Cardiovascular Specialty marketing brochures distributed for Flo-Rester Disposable Internal Vessel Occluder, Flo-Thru Intraluminal Shunt, and Vascular Probe Intravascular Probe contain inaccurate content that does not align with the products Instructions for Use (IFU).
The Surgical Tools and Cardiovascular Specialty marketing brochures distributed for Flo-Rester Disposable Internal Vessel Occluder, Flo-Thru Intraluminal Shunt, and Vascular Probe Intravascular Probe contain inaccurate content that does not align with the products Instructions for Use (IFU).
Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility
The Surgical Tools and Cardiovascular Specialty marketing brochures distributed for Flo-Rester Disposable Internal Vessel Occluder, Flo-Thru Intraluminal Shunt, and Vascular Probe Intravascular Probe contain inaccurate content that does not align with the products Instructions for Use (IFU).
Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility
Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility
Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility
Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.
Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility