🔪

Surgical Instruments

🏥 Medical Devices 7,015 recalls

Medline brand, medical procedure convenience kits, labeled as: 1) LAMINECTOMY CDS-LF, SKU CDS780045N; 2) NEURO CRANIOTOMY CDS #36-RF, SKU CDS780119U; 3) NEURO CRANIOTOMY CDS #36-RF, SKU CDS780119V; 4) CRANIOTOMY CDS, SKU CDS981753K; 5) LAMINECTOMY CDS-LF, SKU CDS981923X; 6) CRANIOTOMY, SKU CDS982719Q; 7) CRANIOTOMY, SKU CDS982719R; 8) CRANIOTOMY, SKU CDS982719S; 9) CRANIOTOMY, SKU CDS982719T; 10) CRANIOTOMY, SKU CDS982719U; 11) CRANIOTOMY, SKU CDS982719V; 12) CRANIOTOMY, SKU CDS982719W; 13) CRANIOTOMY, SKU CDS984074I; 14) SPINE, SKU CDS984943I; 15) SPINE, SKU CDS984943J; 16) KIT NEURO CRANIOTOMY FOR TUMOR, SKU DYKMBNDL144; 17) KIT NEURO CRANIOTOMY FOR TUMOR, SKU DYKMBNDL144A; 18) CRANIOTOMY #IMF 56646-LF, SKU DYNJVB91001; 19) NEURO SPINE #IMF 56655-LF, SKU DYNJVB91004; 20) CRANIOTOMY HARPER PACK-LF, SKU DYNJ0162022G; 21) CRANIOTOMY PACK-LF, SKU DYNJ0530906AP; 22) CRANIOTOMY PACK-LF, SKU DYNJ0530906AQ; 23) LAMINECTOMY PACK-LF, SKU DYNJ17452R; 24) LAMINECTOMY PACK-LF, SKU DYNJ17452S; 25) CRANIOTOMY PACK-LF, SKU DYNJ17455J; 26) CRANIOTOMY PACK-LF, SKU DYNJ17455K; 27) DEPAUL CRANIOTOMY-LF, SKU DYNJ20018P; 28) SPINAL PROCEDURE SC, SKU DYNJ33402K; 29) LAMINECTOMY PACK SC, SKU DYNJ35328J; 30) CRANIOTOMY PACK 9128175-LF, SKU DYNJ35982J; 31) LAMINECTOMY PACK, SKU DYNJ37683J; 32) SPINAL RECONSTRUCTION PACK, SKU DYNJ37689G; 33) CRANIOTOMY PACK, SKU DYNJ37714G; 34) CRANIOTOMY PACK, SKU DYNJ38317G; 35) SPINE PACK-LF, SKU DYNJ38889I; 36) ORTHO SPINE, SKU DYNJ40221A; 37) ORTHO SPINE, SKU DYNJ40221B; 38) ORTHO SPINE, SKU DYNJ40221C; 39) ORTHO SPINE, SKU DYNJ40221D; 40) ORTHO SPINE, SKU DYNJ40221F; 41) CRANIOTOMY PK, SKU DYNJ40388B; 42) CRANIOTOMY-LF, SKU DYNJ43319F; 43) NEURO SPINE PACK 319728, SKU DYNJ44628D; 44) CRANI PACK-LF, SKU DYNJ44805L; 45) LAMINECTOMY PACK-LF, SKU DYNJ44815J; 46) NEURO SPINE PACK, SKU DYNJ45419D; 47) NEURO SPINE PACK, SKU DYNJ45419F; 48) SPINE PACK, SKU DYNJ46317F; 49) CRANI PACK, SKU DYNJ47605B; 50) SCRIPPS SW CRANIOTOMY PACK-LF, SKU DYNJ49653C; 51) PK) NEURO-CRANI, SKU DYNJ53453A; 52) ACDF PACK-LF, SKU DYNJ57136A; 53) PEDI CRANI MAJOR, SKU DYNJ59338C; 54) UMC CRANIOTOMY, SKU DYNJ60497A; 55) UMC CRANIOTOMY PK, SKU DYNJ60497B; 56) CMC NEURO PACK-LF, SKU DYNJ61079C; 57) CMC NEURO PACK-LF, SKU DYNJ61079D; 58) CRANIOTOMY PK, SKU DYNJ65934B; 59) METREX PACK-LF, SKU DYNJ66104A; 60) CRANIOTOMY PACK, SKU DYNJ66419; 61) CRANIOTOMY PACK, SKU DYNJ66419A; 62) CRANIOTOMY PACK RFID, SKU DYNJ66419B; 63) CRANIOTOMY PACK, SKU DYNJ67355A; 64) CRANIOTOMY PACK, SKU DYNJ67355B; 65) OSC LAMINECTOMY PACK, SKU DYNJ67395; 66) SPINE PACK, SKU DYNJ67505F; 67) SPINE PACK, SKU DYNJ67505G; 68) SPINE PACK, SKU DYNJ67652A; 69) NEURO CRANI PACK, SKU DYNJ69752; 70) SPINE PACK, SKU DYNJ69753; 71) SPINE PACK, SKU DYNJ80550A; 72) CRANI PACK, SKU DYNJ80554A; 73) SPINE PACK, SKU DYNJ82105; 74) LAMINECTOMY I-LF, SKU DYNJ83783; 75) ANTERIOR CERVICAL-LF, SKU DYNJ83785; 76) NEURO-CRANI PK-LF, SKU DYNJ85475; 77) CRANIOTOMY PACK, SKU DYNJ85527A; 78) CRANIOTOMY-LF, SKU DYNJ900718D; 79) WMC ANTERIOR CERVICAL-LF, SKU DYNJ902443D; 80) WMC LAMINECTOMY-LF, SKU DYNJ902468J; 81) WMC LAMINECTOMY-LF, SKU DYNJ902468K; 82) WMC LAMINECTOMY-LF, SKU DYNJ902468L; 83) WMC LAMINECTOMY-LF, SKU DYNJ902468M; 84) WMC CRANIOTOMY ADD ON-LF, SKU DYNJ902473L; 85) CRANIOTOMY, SKU DYNJ903224R; 86) CRANIOTOMY, SKU DYNJ903224S; 87) CRANIOTOMY, SKU DYNJ903224T; 88) CRANIOTOMY, SKU DYNJ903224V; 89) DAVIS LAMI, SKU DYNJ903452D; 90) ORTHO INS-SPINE BACK, SKU DYNJ904661D; 91) ORTHO INS-SPINE ACDF, SKU DYNJ904662D; 92) CCOC SPINE EHRLER, SKU DYNJ904681F; 93) KIT) SPINE LH, SKU DYNJ906448B; 94) CRANI, SKU DYNJ906469F; 95) SPINE, SKU DYNJ907030; 96) JVL - MODULE BURR HOLE - CRAN, SKU DYNJ907147A; 97) BGMC MICRODISK-SPINAL, SKU DYNJ907349B; 98) KIT PEDI CRANI LAMI, SKU DYNJ907372A; 99) KIT SPINAL INSTRUMENTATION, SKU DYNJ907376; 100) SPINE, SKU DYNJ907819A; 101) UMC CRANIOTOMY, SKU DYNJ907945; 102)

MEDLINE INDUSTRIES, LP - Northfield

Class I - Dangerous

Surgical Patties & Surgical Strips in specific lots of Custom Convenience Kits were found to contain higher-than-expected levels of endotoxin within the raw material used to produce sterile Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in those finished goods

Nov 6, 2024 Surgical Instruments Nationwide View Details →
Class I - Dangerous

On October 18, 2024, BD has identified through customer complaints and an increase in field failures / replacements of the power supply module (part number 650781, Power Module AC-DC 250W 12/24V M3S4S6) on BD FACSLyric flow cytometers, which has the potential for the instruments to fail to power on and/or stay on. The potential power supply failure on BD FACSLyric Flow Cytometer may lead to risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. The patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In addition, patient specimen loss and/ or delay providing results of testing results. There may be additional risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. In addition, the patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In the Notification Letter sent on November 20, 2024, customers are instructed to continue their normal operation of your BD FACSLyric Flow Cytometer according to the Instructions For Use after immediately inspecting inventory for the specific catalog and power supply serial numbers listed in the notification letter and follow the instructions listed under Actions for Clinical Users.

Nov 1, 2024 Surgical Instruments Nationwide View Details →

Instructions for use update on HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 pertaining to Power-on self-test error code "EA POST".

Dec 4, 2024 Surgical Instruments View Details →

Instructions for use update on HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 pertaining to Power-on self-test error code "EA POST".

Dec 4, 2024 Surgical Instruments View Details →

Due to complaints received for the marker band detaching from the distal tip of the catheter and remaining in the patient as laser catheter was retracted.

Nov 25, 2024 Surgical Instruments Nationwide View Details →

Units were not intended for commercial distribution and inadvertently distributed.

Oct 30, 2023 Surgical Instruments View Details →

BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.

Oct 31, 2024 Surgical Instruments Nationwide View Details →

BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.

Oct 31, 2024 Surgical Instruments Nationwide View Details →

BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.

Oct 31, 2024 Surgical Instruments Nationwide View Details →

BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.

Oct 31, 2024 Surgical Instruments Nationwide View Details →

BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.

Oct 31, 2024 Surgical Instruments Nationwide View Details →

BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.

Oct 31, 2024 Surgical Instruments Nationwide View Details →