The DxA 5000 instructions for use may be confusing regarding the re-use of sample IDs. Beckman Coulter has received complaints where the re-use of sample barcodes were not being detected by the system .
Surgical Instruments
🏥 Medical Devices • 7,015 recalls
Speed Stitch Needle Cassette
ArthroCare
Due to a supplier error where a portion of the raw material was inadvertently mixed. As a result, some needles may be made of commercially pure titanium as opposed to 304 stainless steel which does not meet the specification and could potentially break during use.
Pointts sistema de ellmlnact!;o de verrugas OTC - Brazil Catalog Number: 1001-0280 - Product Usage: indicated for over-the-counter a cryosurgical system for the treatment of skin tags.
OraSure Technologies
Unit boxes not properly sealed
Personelle OTC Skin Tag Remover, Canada Catalog Number: 1001-0314 - Product Usage: indicated for over-the-counter a cryosurgical system for the treatment of skin tags.
OraSure Technologies
Unit boxes not properly sealed
CVS Advanced Wart Remover Kit, Catalog Number: 1001-0407 - Product Usage: indicated for over-the-counter a cryosurgical system for the treatment of common warts and plantar warts.
OraSure Technologies
Unit boxes not properly sealed
lntegra LifeSciences has identified through a number of complaints that the handpiece cable melts following sterilization.
An update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Lactate Dehydrogenase assay on RX Instruments. Additional pipette washes can be implemented as described in the technical bulletin. Interference to the Lactate Dehydrogenase (LOH) reagent would be observed as inconsistencies in Quality Control recovery, which may lead to a delay in running patient samples, or erroneous elevated test results.
An update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Lactate Dehydrogenase assay on RX Instruments. Additional pipette washes can be implemented as described in the technical bulletin. Interference to the Lactate Dehydrogenase (LOH) reagent would be observed as inconsistencies in Quality Control recovery, which may lead to a delay in running patient samples, or erroneous elevated test results.
lntegra LifeSciences has identified through a number of complaints that the handpiece cable melts following sterilization.
lntegra LifeSciences has identified through a number of complaints that the handpiece cable melts following sterilization.
Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results
The Alaris PC unit can display incorrect syringe type and/or syringe sizes. This could result in delays in infusion, under-infusion, or over-infusion.
Ion Catheter Reprocessing Cover, IF1000 (Part Number:490114-02) UDI:00886874116845.
Intuitive Surgical
One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or incorrectly engaged to the ION Fully Articulating Catheter which can result in retention of cleaning chemical or water during reprocessing; may cause failure of the ION Fully Articulating Catheter or ION System.
Medcomp 15.SF X 24CM Titan HD CATHETER (CUFF 19CM FROM TIP) Ref: THD155024
Medical Components, Inc
Kits were packaged with the incorrect size valve peelable introducer. The kits include a 15F instead of the correct 16F valve peelable introducer
ION Reprocessing Accessories Kit, IF1000 (Part Number: 490024-03) UDI:00886874119129. The kit includes an ION Catheter Reprocessing Cover (Part Number: 490114-02).
Intuitive Surgical
One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or incorrectly engaged to the ION Fully Articulating Catheter which can result in retention of cleaning chemical or water during reprocessing; may cause failure of the ION Fully Articulating Catheter or ION System.
A degradation defect was found that could lead to holes and donning tears at the folding area around the cuffs.
A degradation defect was found that could lead to holes and donning tears at the folding area around the cuffs.
A degradation defect was found that could lead to holes and donning tears at the folding area around the cuffs.
A degradation defect was found that could lead to holes and donning tears at the folding area around the cuffs.
A degradation defect was found that could lead to holes and donning tears at the folding area around the cuffs.