🏨

Other Medical Devices

🏥 Medical Devices 15,809 recalls

Class I - Dangerous

For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.

Jan 31, 2025 Other Medical Devices Nationwide View Details →

Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices

Jan 21, 2025 Other Medical Devices Nationwide View Details →

SIGNA MR355, SIGNA MR360, NMRI system

GE Healthcare (China) Co.

Class I - Dangerous

For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.

Jan 31, 2025 Other Medical Devices Nationwide View Details →

138913-01 BD PYXIS MEDBANK MINI CR-2HH-1FH-P 169-114 BD PYXIS MEDBANK MINI CR-2HH-1FH-P 138911-01 BD PYXIS MEDBANK MINI CR-4HH-P 169-115 BD PYXIS MEDBANK MINI CR-4HH-P 139001-01 BD PYXIS MEDBANK MN 200 CR-2HH-1FH-P 139002-01 BD PYXIS MEDBANK MN 200 CR-2HH-1FM-P 139000-01 BD PYXIS MEDBANK MN 200 CR-4HH-P 139020-01 BD PYXIS MEDBANK MN 500 CR-2HH-1FM-2DD-P 139018-01 BD PYXIS MEDBANK MN 500 CR-2HH-5FM-P 139019-01 BD PYXIS MEDBANK MN 500 CR-4HH-1FH-3FM-P 139021-01 BD PYXIS MEDBANK MN 500 CR-4HH-4FM-P 139022-01 BD PYXIS MEDBANK MN 500 CR-6HH-3FM-P 138973-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P 155140-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P 169-82 BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P 138975-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P 155141-01 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P 169-83 BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P 139073-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FH 155142-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FH 169-84 BD PYXIS MEDBANK TWR MN CR-12HH-1FH 138976-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P 155143-01 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P 169-85 BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P 138977-01 BD PYXIS MEDBANK TWR MN CR-14HH-P 155144-01 BD PYXIS MEDBANK TWR MN CR-14HH-P 169-86 BD PYXIS MEDBANK TWR MN CR-14HH-P 138951-01 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM 155145-01 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM 169-87 BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM 139072-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FH 155146-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FH 169-88 BD PYXIS MEDBANK TWR MN CR-2HH-6FH 139085-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FM 155147-01 BD PYXIS MEDBANK TWR MN CR-2HH-6FM 169-89 BD PYXIS MEDBANK TWR MN CR-2HH-6FM 138968-01 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P 155148-01 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P 169-90 BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P 138964-01 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P 155149-01 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P 169-91 BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P 138966-01 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P 155150-01 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P 169-92 BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P 138967-01 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P 155151-01 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P 169-93 BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P 138969-01 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P 155152-01 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P 169-94 BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P 138965-01 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P 155153-01 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P 169-95 BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P 138971-01 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P 155154-01 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P 169-96 BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P 138972-01 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P 155155-01 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P 169-97 BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P 138974-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P 155156-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P 169-98 BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P 138970-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P 155157-01 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P 169-99 BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P

CareFusion 303

Class I - Dangerous

Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms.

Jan 8, 2025 Other Medical Devices Nationwide View Details →

Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms.

Jan 8, 2025 Other Medical Devices Nationwide View Details →
Class I - Dangerous

Beckman Coulter has identified an issue which prevents the DxC 500 AU instrument from providing the correct onboard stability (Open Expiration) dates for assays that were loaded prior to a software upgrade (e.g. 1.3 to 1.4, 1.4 to 1.4.1, etc.). As a result, assays that have expired may still be used for sample processing, leading to inaccurate or erroneous patient test results. The issue only happens on assays loaded before instruments doing upgrades and does not impact new installation. It does not impact the assays loaded after upgrade.

Jan 10, 2025 Other Medical Devices View Details →

155288-01 BD PYXIS MEDBANK MINI 1FH-1FM 169-137 BD PYXIS MEDBANK MINI 1FH-1FM 155311-01 BD PYXIS MEDBANK MINI 1FH-2HH 138921-01 BD PYXIS MEDBANK MINI 1HH-1FH-1HM 155159-01 BD PYXIS MEDBANK MINI 1HH-1FH-1HM 169-101 BD PYXIS MEDBANK MINI 1HH-1FH-1HM 139122-01 BD PYXIS MEDBANK MINI 1HH-1HM-1FM 155160-01 BD PYXIS MEDBANK MINI 1HH-1HM-1FM 169-102 BD PYXIS MEDBANK MINI 1HH-1HM-1FM 138912-01 BD PYXIS MEDBANK MINI 1HH-3HM 155161-01 BD PYXIS MEDBANK MINI 1HH-3HM 169-103 BD PYXIS MEDBANK MINI 1HH-3HM 138902-01 BD PYXIS MEDBANK MINI 2FH 169-104 BD PYXIS MEDBANK MINI 2FH 138903-01 BD PYXIS MEDBANK MINI 2FM 155163-01 BD PYXIS MEDBANK MINI 2FM 169-105 BD PYXIS MEDBANK MINI 2FM 138908-01 BD PYXIS MEDBANK MINI 2HH-1FH 169-106 BD PYXIS MEDBANK MINI 2HH-1FH 138906-01 BD PYXIS MEDBANK MINI 2HH-1FM 155165-01 BD PYXIS MEDBANK MINI 2HH-1FM 169-107 BD PYXIS MEDBANK MINI 2HH-1FM 138907-01 BD PYXIS MEDBANK MINI 2HH-2HM 155166-01 BD PYXIS MEDBANK MINI 2HH-2HM 169-108 BD PYXIS MEDBANK MINI 2HH-2HM 139131-01 BD PYXIS MEDBANK MINI 2HM-1FM 155167-01 BD PYXIS MEDBANK MINI 2HM-1FM 169-109 BD PYXIS MEDBANK MINI 2HM-1FM 139091-01 BD PYXIS MEDBANK MINI 3HH-1FG 155168-01 BD PYXIS MEDBANK MINI 3HH-1FG 169-110 BD PYXIS MEDBANK MINI 3HH-1FG 138909-01 BD PYXIS MEDBANK MINI 3HH-1HM 155169-01 BD PYXIS MEDBANK MINI 3HH-1HM 169-111 BD PYXIS MEDBANK MINI 3HH-1HM 138910-01 BD PYXIS MEDBANK MINI 4HH 155310-01 BD PYXIS MEDBANK MINI 4HH 169-112 BD PYXIS MEDBANK MINI 4HH 138915-01 BD PYXIS MEDBANK MINI 4HM 155171-01 BD PYXIS MEDBANK MINI 4HM 169-113 BD PYXIS MEDBANK MINI 4HM 1147-00 MEDBANK MINI MAIN 138943-01 BD PYXIS MEDBANK TWR MN 10HH-1FH-1FM 155081-01 BD PYXIS MEDBANK TWR MN 10HH-1FH-1FM 169-23 BD PYXIS MEDBANK TWR MN 10HH-1FH-1FM 138955-01 BD PYXIS MEDBANK TWR MN 10HH-2FH 155082-01 BD PYXIS MEDBANK TWR MN 10HH-2FH 169-24 BD PYXIS MEDBANK TWR MN 10HH-2FH 138924-01 BD PYXIS MEDBANK TWR MN 10HH-2FM 155083-01 BD PYXIS MEDBANK TWR MN 10HH-2FM 169-25 BD PYXIS MEDBANK TWR MN 10HH-2FM 138960-01 BD PYXIS MEDBANK TWR MN 12HH-1FF 155084-01 BD PYXIS MEDBANK TWR MN 12HH-1FF 169-26 BD PYXIS MEDBANK TWR MN 12HH-1FF 138961-01 BD PYXIS MEDBANK TWR MN 12HH-1FH 155085-01 BD PYXIS MEDBANK TWR MN 12HH-1FH 169-27 BD PYXIS MEDBANK TWR MN 12HH-1FH 138926-01 BD PYXIS MEDBANK TWR MN 12HH-1FM 155086-01 BD PYXIS MEDBANK TWR MN 12HH-1FM 169-28 BD PYXIS MEDBANK TWR MN 12HH-1FM 138927-01 BD PYXIS MEDBANK TWR MN 14HH 155087-01 BD PYXIS MEDBANK TWR MN 14HH 169-29 BD PYXIS MEDBANK TWR MN 14HH 155293-01 BD PYXIS MEDBANK TWR MN 1FH-10HM-1FM 169-138 BD PYXIS MEDBANK TWR MN 1FH-10HM-1FM 155303-01 BD PYXIS MEDBANK TWR MN 1FM-10HH-1FH 139082-01 BD PYXIS MEDBANK TWR MN 1HH-1HM-2FM-4FF 155088-01 BD PYXIS MEDBANK TWR MN 1HH-1HM-2FM-4FF 169-30 BD PYXIS MEDBANK TWR MN 1HH-1HM-2FM-4FF 155305-01 BD PYXIS MEDBANK TWR MN 2FH-1FM-4HH-2FM 139056-01 BD PYXIS MEDBANK TWR MN 2FH-2HM-4FM 155089-01 BD PYXIS MEDBANK TWR MN 2FH-2HM-4FM 169-31 BD PYXIS MEDBANK TWR MN 2FH-2HM-4FM 155304-01 BD PYXIS MEDBANK TWR MN 2FM-8HH-1FH 139045-01 BD PYXIS MEDBANK TWR MN 2HH-10HM-1FM 155090-01 BD PYXIS MEDBANK TWR MN 2HH-10HM-1FM 169-32 BD PYXIS MEDBANK TWR MN 2HH-10HM-1FM 138932-01 BD PYXIS MEDBANK TWR MN 2HH-1FH-5FM 155091-01 BD PYXIS MEDBANK TWR MN 2HH-1FH-5FM 169-33 BD PYXIS MEDBANK TWR MN 2HH-1FH-5FM 139084-01 BD PYXIS MEDBANK TWR MN 2HH-2FH-4FM 155092-01 BD PYXIS MEDBANK TWR MN 2HH-2FH-4FM 169-34 BD PYXIS MEDBANK TWR MN 2HH-2FH-4FM 138963-01 BD PYXIS MEDBANK TWR MN 2HH-2FM-4FF 155093-01 BD PYXIS MEDBANK TWR MN 2HH-2FM-4FF 169-35 BD PYXIS MEDBANK TWR MN 2HH-2FM-4FF 155294-01 BD PYXIS MEDBANK TWR MN 2HH-4FH-2FM 169-139 BD PYXIS MEDBANK TWR MN 2HH-4FH-2FM 138938-01 BD PYXIS MEDBANK TWR MN 2HH-4HM-4FF 155094-01 BD PYXIS MEDBANK TWR MN 2HH-4HM-4FF 169-36 BD PYXIS MEDBANK TWR MN 2HH-4HM-4FF 139125-01 BD PYXIS MEDBANK TWR MN 2HH-4HM-4FM 155095-01 BD PYXIS MEDBANK TWR MN 2HH-4HM-4FM 169-37 BD PYXIS MEDBANK TWR MN 2HH-4HM-4FM 138919-01 BD PYXIS MEDBANK TWR MN 2HH-6FM 155096-01 BD PYXIS MEDBANK TWR MN 2HH-6FM 169-38 BD PYXIS MEDBANK TWR MN 2HH-6

CareFusion 303

Class I - Dangerous

Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms.

Jan 8, 2025 Other Medical Devices Nationwide View Details →

Automated dispensing device labeling is being updated to strengthen labeling around manual access to medications to address the risk of potential delays in retrieving contents/ accessing medications that may result from power loss or potential automated dispensing device malfunctions.

Jan 10, 2025 Other Medical Devices Nationwide View Details →

Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices

Jan 21, 2025 Other Medical Devices Nationwide View Details →

IntelePACS (Image Fusion Module) - InteleViewer

INTELERAD MEDICAL SYSTEMS INCORPORATED

Class I - Dangerous

A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.

Feb 5, 2025 Other Medical Devices Nationwide View Details →

Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices

Jan 21, 2025 Other Medical Devices Nationwide View Details →

SIGNA MR380, NMRI system, Non-US Only

GE Healthcare (China) Co.

Class I - Dangerous

For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.

Jan 31, 2025 Other Medical Devices Nationwide View Details →

Optima MR360, NMRI system

GE Healthcare (China) Co.

Class I - Dangerous

For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.

Jan 31, 2025 Other Medical Devices Nationwide View Details →