Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
Other Medical Devices
🏥 Medical Devices • 15,809 recalls
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
Engaging titanium base was manufactured 180 degrees around the post from where it should be located in relation to the connection geometry; if the flat is on the wrong side of the post, the crown will not fit onto the abutment in the correct orientation, resulting in restoration delay.
FloPatch FP120, REF: FP120-FOT01-005
Flosonics Medical (R/A 1929803 ONTARIO CORP.)
Flowmeter patch as a human readable expiry date on the outer shipping packaging that reads 2025-06-20 when it should read 2026-06-20.
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
IDM-ARLITE-R. Electrical wheelchair component.
mo-Vis BVBA
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
IDM-MULTI-R. Electrical wheelchair component.
mo-Vis BVBA
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
Potential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series.
Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
TRANSDUCER L12-5 Transducer Probe
Philips Ultrasound
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Medline Kits containing BD SureStep Foley Tray Systems: SKU/Description: CDS984076J GYN ROBOTIC DYNJG001004A LAVH MINOR DYNJG901001B CRANIOTOMY DYNJ908819B KIT CRANIOTOMY (only units from SKU DYNJG001004A LAVH MINOR were distributed)
MEDLINE INDUSTRIES, LP - Northfield
Kits contain SureStep Foley Tray Systems recalled by BD that include incorrect product information insert sheets. If a mislabeled tray is used, patient risk ranges from minor injury to injury that requires medical intervention, such as infection or local/systemic allergic reaction (anaphylaxis).
C6-2 Transducer Probe
Philips Ultrasound
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
L12-5 38mm Transducer Probe
Philips Ultrasound
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
C8-4v Transducer Probe
Philips Ultrasound
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel.
L9-3 Transducer Probe
Philips Ultrasound
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.