Other Medical Devices

πŸ₯ Medical Devices β€’ 15,809 recalls

Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.

Dec 8, 2025 Other Medical Devices Nationwide View Details β†’

Allura Xper FD10; Model Number: 722003;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Class I - Dangerous

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

Dec 15, 2025 Other Medical Devices Nationwide View Details β†’

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

Dec 16, 2025 Other Medical Devices Nationwide View Details β†’

Allura Xper FD10/10; Model Number: 722011;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Class I - Dangerous

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

Dec 15, 2025 Other Medical Devices Nationwide View Details β†’

Allura Xper FD20/20; Model Number: 722038;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Class I - Dangerous

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

Dec 15, 2025 Other Medical Devices Nationwide View Details β†’

Allura Xper FD10/10; Model Number: 722005;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Class I - Dangerous

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

Dec 15, 2025 Other Medical Devices Nationwide View Details β†’

Allura Xper FD20; Model Number: 722012;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Class I - Dangerous

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

Dec 15, 2025 Other Medical Devices Nationwide View Details β†’

Medline medical procedure convenience kits labeled as: 1) LAPAROSCOPY CDS-LF, REF CDS860235L; 2) GENERAL LAPAROSCOPIC, REF CDS930223M; 3) LAP CHOLE CDS, REF CDS983678K; 4) LAPAROSCOPY PELV CDS, REF CDS983679K; 5) LAP CHOLE CDS-LF, REF CDS983752M; 6) LAP CHOLE CDS-LF, REF CDS984016B; 7) LAP CHOLE KIT, REF CDS984020P; 8) LAP CHOLE-LF, REF CDS984175J; 9) LAP COLON CDS, REF CDS984849N; 10) GENERAL LAPAROSCOPY CDS, REF CDS985328M; 11) LAP CHOLE, REF DYNJQ9044O; 12) GENERAL LAPAROSCOPY V, REF DYNJS2032F; 13) LAP CHOLE PACK, REF DYNJ26209R; 14) LAP CHOLE PACK, REF DYNJ26209S; 15) GENERAL LAPAROSCOPY PACK, REF DYNJ37459D; 16) DR. BROWN LAP CHOLE PLUS SETUP, REF DYNJ42578G; 17) LAPAROSCOPY GENERAL PACK, REF DYNJ45083G; 18) LAP CHOLE PACK, REF DYNJ51785D; 19) LAPAROSCOPY PACK, REF DYNJ61652A; 20) ENDOSCOPY, REF DYNJ62133A; 21) LAP CHOLE PACK, REF DYNJ66285A; 22) LAPAROSCOPY PACK, REF DYNJ66289B; 23) LAPAROSCOPIC CHOLECYSTECTO, REF DYNJ67296D; 24) LAPAROSCOPY PACK, REF DYNJ68187A; 25) LAP OB PACK, REF DYNJ68741A; 26) LAP CHOLE PACK-LF, REF DYNJ81408A; 27) LAP CHOLE PACK-LF, REF DYNJ81408B; 28) LAP CHOLE PACK, REF DYNJ85318; 29) GENERAL LAPAROSCOPY PACK, REF DYNJ89677; 30) LEX GENERAL LAPAROSCOPY, REF DYNJ900969J; 31) LAP CHOLE, REF DYNJ901126G; 32) BARIATRIC KIT, REF DYNJ901145I; 33) GENERAL LAPAROSCOPY SOMC-LF, REF DYNJ901164M; 34) LAP GASTRIC SLEEVE, REF DYNJ901504L; 35) GENERAL LAP GASTRIC, REF DYNJ901827X; 36) GENERAL LAPAROSCOPY, REF DYNJ901847N; 37) LOU LAP SIGMOID COLON, REF DYNJ901848L; 38) LAP CHOLE, REF DYNJ901870J; 39) LEX BARIATRIC, REF DYNJ902033L; 40) BARIATRIC, REF DYNJ902515K; 41) LAP CHOLE CDS, REF DYNJ902565K; 42) LAP CHOLE, REF DYNJ902995I; 43) LAP CHOLE, REF DYNJ903163L; 44) LAP APPY, REF DYNJ903164L; 45) GENERAL SURGERY LAPAROSCOPY, REF DYNJ903733C; 46) LAP CHOLE, REF DYNJ904693D; 47) BARIATRIC, REF DYNJ904891N; 48) LAPAROSCOPY, REF DYNJ905071F; 49) KIT GENERAL LAPAROSCOPY MHS, REF DYNJ905390F; 50) KIT GENERAL LAPAROSCOPY MHS, REF DYNJ905390G; 51) KIT GENERAL LAPAROSCOPY WEILER, REF DYNJ905407D; 52) KIT GENERAL LAP WAKEFIELD, REF DYNJ905413A; 53) KIT GENERAL LAP WAKEFIELD, REF DYNJ905413B; 54) LAP BARIATRIC, REF DYNJ905733K; 55) KIT LAPAROSCOPIC GASTRIC B, REF DYNJ905856A; 56) LAP CHOLE, REF DYNJ906277D; 57) LAP CHOLE RFD, REF DYNJ906302D; 58) KIT LAP COLON RFD, REF DYNJ906315D; 59) GASTRIC SLEEVE, REF DYNJ906355D; 60) PROSTATECTOMY LAP ROBOTIC, REF DYNJ906805C; 61) LAP ROBOTIC SURG ONCOLOGY, REF DYNJ906824C; 62) GEN BARIATRIC, REF DYNJ907639D; 63) LAP CHOLE, REF DYNJ908415B; 64) LAP CHOLE, REF DYNJ908415C; 65) LAPAROSCOPIC, REF DYNJ908569B; 66) ROBOTIC GASTRIC LAP SLEEVE-LF, REF DYNJ909119G; 67) LAPAROSCOPIC GASTRIC ADD ON, REF DYNJ909628A; 68) BARIATRIC CENTRASTATE, REF DYNJ909817A; 69) LAP GENERAL THEDA, REF DYNJ910050; 70) LAPAROSCOPIC CHOLE PACK, REF DYNJ910139A; 71) ROBOTIC THORACIC, REF DYNJ910386; 72) BARIATRIC, REF DYNJ911017.

Medline Industries

Class I - Dangerous

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

Dec 16, 2025 Other Medical Devices Nationwide View Details β†’
Class I - Dangerous

There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.

Dec 19, 2025 Other Medical Devices Nationwide View Details β†’

Allura Xper FD20/10; Model Number: 722029;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Class I - Dangerous

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

Dec 15, 2025 Other Medical Devices Nationwide View Details β†’

SMV DSI, System, Tomography, Computed, Emission

GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING

Class I - Dangerous

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.

Dec 10, 2025 Other Medical Devices View Details β†’

Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.

Dec 12, 2025 Other Medical Devices Nationwide View Details β†’

Vision RT is informing customers of an omission identified in the Instructions for Use for AlignRT InBore systems. Information identifying the lasers and their characteristics was omitted from the Instructions for Use for the subject system as required by 21 CFR 1040,10(h), via conformance with recognized consensus standard IEC 60825-1 Ed. 3 (FDA Laser Notice 56).

Aug 27, 2025 Other Medical Devices View Details β†’

Due a software issue that may result in equipment not receiving timely updates of patient, medication, or related information from enterprise systems.

Dec 3, 2025 Other Medical Devices Nationwide View Details β†’

Firm received multiple complaints regarding tubing disconnecting from oxygen mask. If issue occurs during use, may lead to delay in patient care and patient may experience shortness of breath leading to hypoxia, which may require medical intervention to prevent further respiratory deterioration.

Nov 21, 2025 Other Medical Devices Nationwide View Details β†’

Beckman Coulter has determined that, due to a potential communication issue, the analyzers listed above may not successfully restart following a standard shutdown, even after multiple attempts. If this issue occurs, the instrument will be unable to complete initialization and transition to Ready mode and the instrument will not process patient samples, QCs or calibrators. This causes delayed results.

Nov 26, 2025 Other Medical Devices Nationwide View Details β†’