SARS-CoV-2 Antigen Assay test may generate false positive results which could result in incorrect diagnosis .
Infusion Pumps
🏥 Medical Devices • 4,350 recalls
Medtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a higher frequency of neurological adverse events and mortality associated with the system when compared to other commercially available left ventricular assist devices, as well as complaints that the internal pump may delay or fail to restart.
Alaris Infusion Pump Module, Model 8100, serviced/repaired by Biomedical Equipment Service Co. (BMES) with an impacted bezel repair part TIPA-8100-4410
BioMedical Equipment Service Co (BMES)
Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fluids.
Baxter Healthcare
System errors were noted following changes the customers implemented to the configuration of their network and server systems. Gateway Server System performance anomalies resulted in excessive Spectrum IQ pump connectivity errors.
Medtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a higher frequency of neurological adverse events and mortality associated with the system when compared to other commercially available left ventricular assist devices, as well as complaints that the internal pump may delay or fail to restart.
Potential for a manufacturing assembly error-may result in a non-functional audio indicator of device activation or a non-functional device, may cause unintended thermal burn, hemorrhage, or tissue damage and a delay of treatment while an alternate device is located
Prime TC-transport chair intended for medical purposes to assist a person in performing an activity that the person would find difficult to do or be unable to do Model Number: 1460, Part #1460000000¿¿¿¿¿¿¿
Stryker Medical Division of Stryker
Medical devices intended for use in a healthcare facility were distributed to residential customers and may be in use in a home environment. The product s labeling accurately identifies the correct use environment, but home users may not be aware of or fully understand these instructions intended for healthcare professionals
REF 8L92-42, Abbott Activated Alanine Aminotransferase, for use with Architect, IVD, CE, UDI: (01) 00380740161866
Abbott Laboratories
There is a potential for >10% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results
Baxter Healthcare Corporation has identified that the AMIA Sharesource User Guide that is used with the AMIA Automated PD System cyclers incorrectly describes the function of the UF (Ultrafiltration) Limit setting as the ending criteria of the cyclers Extra Last Drain Mode. In some locations in the guide, it states that the UF volume achieved must exceed the programmed UF Limit before the cycler will end the Extra Last Drain, allowing the patient to end therapy or proceed to their last fill.
REF: 08P1820, Abbott Alinity c Activated Alanine Aminotransferase reagent Kit, Alinity c Activated Alanine Aminotransferase Reagent Kit,, IVD, CE, UDI: (01) 00380740132569
Abbott Laboratories
There is a potential for 15% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results.
REF 8L92-21, Abbott Activated Alanine Aminotransferase, for use with Architect, IVD, CE, UDI: (01) 00380740017866
Abbott Laboratories
There is a potential for >10% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results
REF 8L92-22, Abbott Activated Alanine aminotransferase, for use with Architect, IVD, CE, UDI: (01) 00380740161859
Abbott Laboratories
There is a potential for >10% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results
There is a potential for the Anti-free flow clip of the administration set to be inverted which may cause unintended free flow if the user fails to follow on screen prompt to close the roller clamp prior to set removal from the pump.
The seating system unexpectedly detached, which resulted in the seating system falling to the ground, causing injury.
The seating system unexpectedly detached, which resulted in the seating system falling to the ground, causing injury.
Latch mechanism on the Lift-Off foot section of the Affinity Four Birthing Bed is damaged, it could potentially cause the installed foot section to be improperly engaged onto the bed, could result in injury to the user due to a fall.
Exposed to freezing temperatures due to refrigerated truck malfunctions
Not properly sterilized, because the outer pouch seal on the Tyvek header was sealed closely to the foil pouch.
NEFA CAL (Non-Esterified Fatty Acids)- IVD Standard for measurement of NEFA Catalog Number: FA115
Randox Laboratories
Standard packed within NEFA kit has been assigned incorrectly. Calibrating with the incorrect value will cause a positive shift of up to +8% with both quality controls and patient samples
Exposed to freezing temperatures due to refrigerated truck malfunctions