Instructions for use were updated.
Infusion Pumps
🏥 Medical Devices • 4,350 recalls
C-Peptide ELISA : IVD measurement of C-Peptide in serum, plasma (EDTA-, heparin- or citrate plasma) and urine. Catalog # EIA-1293
DRG International
ODs of the standards too low(< 0.6 ODmax) may cause invalid runs, delayed patient results are possible
Potential for a loose cylindrical pin on the Transport Guard that may result in the jamming of the locking mechanism causing the Transport Guard to not open or, if the locking pin is fully disengaged from the assembly, a full disassembly of the Transport Guard may occur, resulting in a delay of support if a product exchange is necessary.
This calibrator lot did not meet acceptance criteria during ongoing stability testing. When a calibration is performed with this lot, controls may exceed the specified range in the Instructions for Use (IFU).
An unreleased Version 4 Protocol Library was installed on Nimbus PainPro (H) infusion pumps in error that may impact patient safety if used on patients.
Shelf-life for the 8244 Pneumonia (33 Targets) Control Panel determined to be 6 months instead of 18 months.
Liko Universal Twinbar 670 QRH, an accessory that can be used together with most Liko lifts, model numbers 3156087 and P3156087.
Baxter Healthcare
An incorrect center bolt was used in the assembly of the product which could result in a patient fall during use.
There is a potential that the DNA sequence analyzer may short circuit shortly after disconnection and reconnection of circuit board or imaging module due to an insufficient reconnection of the ribbon cable. The short circuit results in the device being inoperable.
New or replacement insulin pumps are not pre-programmed with basal rates or other verified settings (bolus wizard settings, sensor settings, etc.), which must be set up and saved on pumps prior to use. Patients who do not realize this will receive no basal insulin, which can lead to insulin under-delivery. Insulin under-delivery can cause severe hyperglycemia, which may lead to diabetic ketoacidosis. Current device labeling does not clearly identify that these insulin pumps are capable of functioning with the default basal setting of 0.0 units/hr for all 24 hours and that no error message will appear if the default basal setting remains active. Furthermore, the labeling for the affected devices does not highlight that saving the basal settings requires multiple steps and failure to complete all steps will not save the new basal settings, nor that the steps required to set and save a basal profile and other parameters which are used as part of the bolus calculator function are different.
Mycoplasma genitalium Control Panel (Inactivated Pellet)
Microbiologics
Distributed product did not undergo proper release testing.
Sorin Centrifugal Pump System, CP5, Drive Unit (Catalogue Item #60-01-04, UDI: 04033817900948) used in an HLM S5/C5 System configuration. A cardiopulmonary bypass speed control device.
LivaNova Deutschland GmbH
One ferrite ring expected to be installed on centrifugal pump system power cable may not be present.
Failed stability specifications for reagent kits that may result in false negative test results.
ProteusPLUS and ProteusONE- designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation Proteus 235
Ion Beam Applications S.A.
When resuming from a beam pause, the Proton Therapy System does not verify whether the beam range has not been manually modified during the pause and is still the prescribed one. Therefore, if an operator manually modified the range during a pause, there is a risk that a portion of the treatment beam after the resume is delivered with an error in range
NIM TRIVANTAGE EMG Endotracheal Tube
Medtronic Xomed
There is potential for multiple issues due to a manufacturing nonconformity.
Outer labelling (HEMASHIELD GOLD Knitted Microvel Double Velour Bifurcated Vascular Graft, D: 16 x 8 mm; L: 40 cm) does not correspond to the (inner) sterile packaging labelling (HEMASHIELD GOLD Knitted Microvel Double Velour Straight Vascular Graft, D: 7 mm L: 60 cm). Mislabeled product could cause a disruption of the vascular anastomosis in the long term, creating a pseudo-aneurysm
KWIK-STIK, Quality control kit for culture media, sold as: KWIK-STIK (single-pack) Enterococcus faecalis derived from ATCC 51299
Microbiologics
This strain should be resistant to high levels of Gentamicin and be resistant to Gentamicin Synergy. A specific lot was found susceptible to Gentamicin Synergy during in-house testing.
Real-time shelf life testing failed at 24 months
The Lactobacillus acidophilus negative control (item 0243 single KWIK-STIK) contained in the 8164 Cepheid Xpert GBS QC Set and 8165 Cepheid Smart GBS QC Set was found to test positive for Group B Streptococcus (GBS) in some cases. This causes a failure when used as a quality control for the Cepheid Xpert GBS and Cepheid Smart GBS systems. The failure of the quality control test causes a disruption for the testing laboratory which may result in repeated testing or the need for alternative test methods. This recall is applicable to the 8164 and 8165 sets only.
da Vinci Xi and X 12-8 mm Conductive Cannula Reducer, REF: 470381-11
Intuitive Surgical
Conductive Cannula Reducer metal tip may get dislodged from the plastic shaft. If unnoticed, additional surgery may be required to remove the metal tip. If the procedure proceeds without the metal tip, there is potential for energy leakage, which could cause tissue damage. If the metal tip remains in the patient, adhesions may form, which may lead to bowel obstruction and chronic abdominal pain.
Helix Elite" Molecular Standards: Inactivated Standard. Inactivated Neisseria gonorrhoeae Strain GL0071.
Microbiologics
Real-time shelf life testing failed at 24 months