A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
Fluid ingress that can cause a loss of electrical funtion and failure of the Set ID Sensory, resulting in the prevention of infusion or halt of infusion.
Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positivesBio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positivesBio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positives.There is no performance issue with the calibrator lot# 53727. They are included in the recall only because the calibrators are lot matched to the impacted reagent pack.
Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positives.
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
A limited number of remediated Philips DreamStation units may experience communication issues when connecting to the cloud-based care management application. Should this be the case, this may affect the ability to download prescription settings to the device or the device may have incorrect prescription settings for the patient.
The Antibody (R2 reagent) is showing a positive bias compared to previous batches on patient samples. Quality Controls do not show the same bias and will be within range
There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB inflation during therapy or a reduction in the degree of augmentation provided to the patient during therapy.
There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB inflation during therapy or a reduction in the degree of augmentation provided to the patient during therapy.
There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB inflation during therapy or a reduction in the degree of augmentation provided to the patient during therapy.
There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB inflation during therapy or a reduction in the degree of augmentation provided to the patient during therapy.
There have been reported failures of the high pressure helium regulator, which may cause a helium leak in the Cardiosave Hospital Cart. In instance of helium regulator failure, a Pump Console s internal reservoir of helium will not be replenished when docked into an impacted Hospital Cart. This may result in an insufficient amount of helium within the internal reservoir.
Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.
New warnings and precautionary statements on product labelling: Reagent 2 (R2) of BIOPHEN Protein C 5 contains the chemical substance - Cesium Chloride which changed classification and has become hazardous according to the CLP regulation ((EC) No 1272/2008) is suspected of damaging fertility