šŸ’§

Infusion Pumps

šŸ„ Medical Devices • 4,350 recalls

Class I - Dangerous

GE HealthCare has become aware that certain Vivid ultrasound systems cannot boot up in a timely fashion. If this occurs, it can delay availability of the device in time-critical emergency situations.

May 30, 2023 Infusion Pumps View Details →

Potential for lot number discrepancies between the test strip packaging and the barcode label applied to the test strip that is read by the instrument for LOT 123909 of NeuMoDx SARS-CoV-2 Test Strips (REF 300800). The instrument registers the non-human readable text as lot number 123910 from the barcode label.

May 15, 2023 Infusion Pumps Nationwide View Details →

When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.

May 10, 2023 Infusion Pumps View Details →

When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.

May 10, 2023 Infusion Pumps View Details →

Device may experience malfunctions due to misalignment of the check valve in the manifold. If the device fails, there is a risk of nitric oxide leak, oxygen gas leak, or earlier cylinder gas changeover than expected or not at all. In a worst case scenario, there may be an interruption in therapy which could lead to oxygen desaturation of the patient and associated complications.

May 3, 2023 Infusion Pumps Nationwide View Details →

IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESSĀæ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESSĀæ Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time

Apr 13, 2023 Infusion Pumps Nationwide View Details →

IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESSĀæ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESSĀæ Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time

Apr 13, 2023 Infusion Pumps Nationwide View Details →

IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESSĀæ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESSĀæ Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time

Apr 13, 2023 Infusion Pumps Nationwide View Details →

IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESSĀæ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESSĀæ Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time

Apr 13, 2023 Infusion Pumps Nationwide View Details →

IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESSĀæ Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESSĀæ Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time

Apr 13, 2023 Infusion Pumps Nationwide View Details →

There is a potential for a disruption in one or more of the triple layer membranes of the ventricular assist device. Recall occurred in 2018, and was retrospectively reported.

May 18, 2018 Infusion Pumps View Details →