πŸ’§

Infusion Pumps

πŸ₯ Medical Devices β€’ 4,350 recalls

Foam gasket on the Evidence MultiSTAT chip heater assembly may not perform as expected. Foam that is not functioning as expected may cause light to leak into the Charged Coupled Device (CCD) Camera during imaging. This will affect the testing and may delay in reporting results.

Apr 5, 2024 Infusion Pumps Nationwide View Details β†’

Due to receiving information regarding difficulties with the insertion of a setscrew inside a screw head. Issues involved, either: 1. The setscrew is not tightened to torque 2. The setscrew is tightened to torque but position askew

Mar 15, 2024 Infusion Pumps Nationwide View Details β†’

During normal use, the mobile app version 2.7 may crash and be automatically relaunched by the iOS operating system. This cycle intermittently repeats, which leads to excessive Bluetooth communication that may result in pump battery drain and may lead to the pump shutting down sooner than typically expected. Pump shutdown will cause insulin delivery to suspend, which could lead to an under-delivery of insulin and may result in hyperglycemia, including severe hyperglycemia.

Mar 5, 2024 Infusion Pumps Nationwide View Details β†’

Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatment

Mar 22, 2024 Infusion Pumps Nationwide View Details β†’

Specific lots of Neria Soft Infusion Sets, which have been shipped into the US Market, have been shipped with the incorrect Instruction for Use (IFU) within the final pack. The IFU included is the IFU registered for the Rest of World (ROW) countries. The difference between the US IFU and the ROW IFU is the addition of an intended use section in the ROW IFU.

Mar 5, 2024 Infusion Pumps Nationwide View Details β†’

Potential for falsely elevated Chol_2, LDLC, and Trig_2 results on the Atellica CH and Atellica CI analyzers when the previous result in the cuvette was Iron3. Results in a positive bias ranging from 2-16% -impacts calibrator, quality control (QC), and patient results.

Mar 13, 2024 Infusion Pumps Nationwide View Details β†’

Inability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure to exchange the device for another leading to leading to a cascade of events which may include additional intervention to close the puncture site, bleeding, leakage of gastric contents, and aspiration

Feb 20, 2024 Infusion Pumps Nationwide View Details β†’

Inability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure to exchange the device for another leading to leading to a cascade of events which may include additional intervention to close the puncture site, bleeding, leakage of gastric contents, and aspiration

Feb 20, 2024 Infusion Pumps Nationwide View Details β†’

Inability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure to exchange the device for another leading to leading to a cascade of events which may include additional intervention to close the puncture site, bleeding, leakage of gastric contents, and aspiration

Feb 20, 2024 Infusion Pumps Nationwide View Details β†’