Implants & Prosthetics

πŸ₯ Medical Devices β€’ 4,589 recalls

Neurovascular embolization device. The Optima Coil System is a series specialized coils that are inserted into the vasculature under angiographic visualization to embolize intracranial aneurysms and other vascular anomalies. The system consists of an embolization coil implant comprised of platinum/tungsten, affixed to a delivery pusher to facilitate insertion into the hub of a microcatheter. The system is available in various shapes, lengths and sizes. The devices are to be placed into aneurysms to create blood stasis, reducing flow into the aneurysm and thrombosing the aneurysm. Upon positioning coils into the aneurysm, the coils are thermally detached from the delivery pusher in serial manner until the aneurysm is occluded. There are a total of 208 Optima Coil models. Optima Coil Model Numbers and Product description: OPTI0101CSS10 Optima Coil 1mm x 1cm Complex-10 SuperSoft OPTI0101CSS10 Optima Coil 1mm x 1cm Complex-10 SuperSoft OPTI0101HSS10 Optima Coil 1mm x 1cm Helical-10 SuperSoft OPTI0101HSS10 Optima Coil 1mm x 1cm Helical-10 SuperSoft OPTI0101HSS10 Optima Coil 1mm x 1cm Helical-10 SuperSoft OPTI0101HSS10 Optima Coil 1mm x 1cm Helical-10 SuperSoft OPTI0101HSS10 Optima Coil 1mm x 1cm Helical-10 SuperSoft OPTI0101HSS10 Optima Coil 1mm x 1cm Helical-10 SuperSoft OPTI0101HSS10 Optima Coil 1mm x 1cm Helical-10 SuperSoft OPTI0101HSS10 Optima Coil 1mm x 1cm Helical-10 SuperSoft OPTI0101HSS10 Optima Coil 1mm x 1cm Helical-10 SuperSoft OPTI0101HSS10 Optima Coil 1mm x 1cm Helical-10 SuperSoft OPTI0101HSS10 Optima Coil 1mm x 1cm Helical-10 SuperSoft OPTI0101HSS10 Optima Coil 1mm x 1cm Helical-10 SuperSoft OPTI0101HSS10 Optima Coil 1mm x 1cm Helical-10 SuperSoft OPTI0101HSS10 Optima Coil 1mm x 1cm Helical-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Optima Coil 1mm x 2cm Complex-10 SuperSoft OPTI0102CSS10 Opti

BALT USA

Class I - Dangerous

Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.

Jan 3, 2024 Implants & Prosthetics Nationwide View Details β†’

The Prestige Coil System is a product family of embolic coils with associated delivery system components. These devices are angiographically delivered through the vasculature to embolize peripheral vascular abnormalities. The devices are to be placed to create blood stasis, reducing flow into the anomaly, and thrombosing the target site. The Prestige Coil System is comprised of the implant coil and a pusher assembly. There are a total of 81 Prestige coil models. Vascular Embolization Device Prestige Coil Model Numbers and Product description: PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack PRES1137COM18 Prestige Coil 11mm x 37cm Complex-18 PRES0202CXPPL Prestige Coil 2mm x 2cm Prestige Plus Complex Xtra Pac PRES0203CXPPL Prestige Coil 2mm x 3cm Prestige Plus Complex Xtra Pac PRES0410CPKPL Prestige Coil 4mm x 10cm Prestige Plus Complex Pack PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack PRES0315HXPPL Prestige Coil 3mm x 15cm Prestige Plus Helical Xtra Pac PRES0420HPKPL Prestige Coil 4mm x 20cm Prestige Plus Helical Pack PRES0520HPKPL Prestige Coil 5mm x 20cm Prestige Plus Helical Pack PRES0620HPKPL Prestige Coil 6mm x 20cm Prestige Plus Helical Pack PRES0720CPKPL Prestige Coil 7mm x 20cm Prestige Plus Complex Pack PRES0930CPKPL Prestige Coil 9mm x 30cm Prestige Plus Complex Pack PRES0101HXPPL Prestige Coil 1mm x 1cm Prestige Plus Helical Xtra Pac PRES0153HXPPL Prestige Coil 1.5mm x 3cm Prestige Plus Helical Xtra Pac PRES0720HPKPL Prestige Coil 7mm x 20cm Prestige Plus Helical Pack PRES0830HPKPL Prestige Coil 8mm x 30cm Prestige Plus Helical Pack PRES0930HPKPL Prestige Coil 9mm x 30cm Prestige Plus Helical Pack PRES1040HPKPL Prestige Coil 10mm x 40cm Prestige Plus Helical Pack PRES1240COM18 Prestige Coil 12mm x 40cm Complex-18 PRES0204CXPPL Prestige Coil 2mm x 4cm Prestige Plus Complex Xtra Pac PRES0206CXPPL Prestige Coil 2mm x 6cm Prestige Plus Complex Xtra Pac PRES0310CXPPL Prestige Coil 3mm x 10cm Prestige Plus Complex Xtra Pac PRES0620CPKPL Prestige Coil 6mm x 20cm Prestige Plus Complex Pack PRES0830CPKPL Prestige Coil 8mm x 30cm Prestige Plus Complex Pack PRES0210CXP10 Prestige Coil 2mm x 10cm Complex-10 Xtra Pack PRES0258CXP10 Prestige Coil 2.5mm x 8cm Complex-10 Xtra Pack PRES0417CPK10 Prestige Coil 4mm x 17cm Complex-10 Xtra Pack PRES0520CPK10 Prestige Coil 5mm x 20cm Complex-10 Xtra Pack PRES0615CPK10 Prestige Coil 6mm x 15cm Complex-10 Xtra Pack PRES0103CXPPL Prestige Coil 1mm x 3cm Prestige Plus Complex Xtra Pac PRES0152CXPPL Prestige Coil 1.5mm x 2cm Prestige Plus Complex Xtra Pac PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack PRES2525HXPPL Prestige Coil 2.5mm x 2.5cm Prestige Plus Helical Xtra Pac PRES1040CPKPL Prestige Coil 10mm x 40cm Prestige Plus Complex Pack PRES0307CXPPL Prestige Coil 3mm x 7cm Prestige Plus Complex Xtra Pac PRES0517CPKPL Prestige Coil 5mm x 17cm Prestige Plus Complex Pack PRES0620CPKPL Prestige Coil 6mm x 20cm Prestige Plus Complex Pack PRES0152CXPPL Prestige Coil 1.5mm x 2cm Prestige Plus Complex Xtra Pac PRES0413CRG10 Prestige Coil 4mm x 13cm Complex 10 Regular PRES0153CXPPL Prestige Coil 1.5mm x 3cm Prestige Plus Complex Xtra Pac PRES0255CXPPL Prestige Coil 2.5mm x 5cm Prestige Plus Complex Xtra Pac PRES0358CXPPL Prestige Coil 3.5mm x 8cm Prestige Plus Complex Xtra Pac PRES0410CPKPL Prestige Coil 4mm x 10cm Prestige Plus Complex Pack PRES1040CPKPL Prestige Coil 10mm x 40cm Prestige Plus Complex Pack PRES0420HPKPL Prestige Coil 4mm x 20cm Prestige Plus Helical Pack PRES0520HPKPL Prestige Coil 5mm x 20cm Prestige Plus Helical Pack PRES0204HXP10 Prestige Coil 2mm x 4cm Helical-10 Xtra Pack PRES0206HXP10 Prestige Coil 2mm x 6cm Helical-10 Xtra Pack PRES1447COM18 Prestige Coil 14mm x 47cm Complex-18 PRES0204HXP10 Prestige Coil 2mm x 4cm Helical-10 Xtra Pack PRES0306HPK10 Prestige Coil 3mm x 6cm Helical-10 Pack PRES0310HPK10 Prestige Coil 3mm x 10cm Helical-10 Pack PRES0410HPK10 Prestige Coil 4mm x 10cm Helical-10 Pack PRES0930H

BALT USA

Class I - Dangerous

Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.

Jan 3, 2024 Implants & Prosthetics Nationwide View Details β†’

Procedural convenience kits and trays, cardiac, labeled as: a) EP LAB BI V IMPLANT PACK, kit number AGBI64J; b) LAB DEVICE IMPLANT, kit number AGDI40J; c) LAB DEVICE IMPLANT, kit number AGDI40K; d) MINOR VASCULAR, kit number AGMV26O; e) OPEN HEART ACCESSORY, kit number AGOH98Q; f) OPEN HEART ACCESSORY, kit number AGOH98R; g) OPEN HEART ACCESSORY, kit number AGOH98S; h) PERIPHERAL VASCULAR, kit number AGPV28O; i) PERIPHERAL VASCULAR, kit number AGPV28P; j) MINOR VASCULAR, kit number AHMV26J; k) MINOR VASCULAR, kit number AHMV26K; l) HEART PACK - 205947, kit number ANCV78AX; m) HEART PACK - 205947 , kit number ANCV78AX1; n) HEART PACK - 205947 Post Open-Heart, kit number ANHK74N; o) OPEN HEART EAST PACK, kit number BHOH43; p) 74OFF PUMP CABG PACK, kit number BHOP48; q) PACEMAKER PACK, kit number BHPM60; r) FORBES EP LAB DEVICE IMPLANT PK, kit number FBDI42K; s) T AND A PACK AGH ASC, kit number FBDI42L; t) T AND A PACK AGH ASC, kit number FBDI42M; u) THORACOSCOPY PACK, kit number FBTO60G; v) OPEN HEART ACCESSORY PACK, kit number JROH00O; w) OPEN HEART ACCESSORY PACK, kit number JROH98O; x) ADULT OPEN HEART PACK, kit number LLOH52; y) ADULT OPEN HEART PACK, kit number LLOP25; z) PACEMAKER PACK, kit number LLPP14; aa) HEART PACK, kit number LMCB06AD; bb) CARDIAC HEART BUNDLE NS, kit number LMCH99O; cc) OPEN HEAART PACK, kit number LMOH50AD; dd) OPEN HEAART PACK, kit number LMOH50AD-01; ee) OPEN HEAART PACK, kit number LMOH50AD-02; ff) OPEN HEAART PACK, kit number LMOH50AE; gg) CV PACK, kit number MMCV98V; hh) CV PACK, kit number MMCV98V-03; ii) CV PACK, kit number MMCV98V-04; jj) KIT OPEN HEART ADULT, kit number MMOH64J; kk) PACEMAKER PACK, kit number MMPM27E; ll) KIT, THORACOTOMY UNIVERSAL, kit number MMTR12F; mm) ANT CERV FUSION PACK, kit number NCAC96; nn) OPEN HEART- ACCESSOR, kit number NCOH11; oo) VASCULAR SAH, kit number SAVA91G; pp) VASCULAR SAH, kit number SAVA91H; qq) VASCULAR SAH, kit number SAVA91I; rr) VASCULAR SAH, kit number SAVA91J; ss) OPEN HEART BRINGBACK SJH, kit number SJOH17I; tt) OPEN HEART CV SJH, kit number SJOH28L; uu) PEDIATRIC HEART SJH, kit number SJPH34K; vv) VASC OPEN CVOR SJH, kit number SJVO56A; ww) VASC OPEN CVOR SJH, kit number SJVO56B; xx) MAJOR CV TRAY, kit number SVMJ20L; yy) MAJOR CV TRAY, kit number SVMJ20M; zz) MINOR CV TRAYΒΏ, kit number SVMN60J; aaa) OPEN HEART COMPONENT PACK, kit number SVOH38Q; bbb) CV B PACK, kit number TNCB10AK; ccc) CARDIOVASCULAR SUPPLY (PS 907285), kit number UICD62AM

American Contract Systems

Class I - Dangerous

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

Nov 29, 2023 Implants & Prosthetics Nationwide View Details β†’