The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
Trividia Health
The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
Manufacturing defect causing LCD display issues with missing or partial segments (ghosting) that can result in misreading glucose values, potentially leading to inappropriate treatment decisions for diabetes management.
One (1) TRUE METRIX AIR blood glucose meter distributed in the United States was packaged into a TRUE METRIX blood glucose meter kit and has an incorrect factory-set unit of measure; the meter displays glucose results in mmol/L rather than mg/dL.
Product gives incorrect low blood glucose levels.
Product gives incorrect low blood glucose levels.
Product gives incorrect low blood glucose levels.
Product gives incorrect low blood glucose levels.