Class I - Dangerous
Heart valves were released and distributed before all required testing was complete, so it cannot be confirmed that they meet all release specifications.
Heart valves were released and distributed before all required testing was complete, so it cannot be confirmed that they meet all release specifications.
Four issues were uncovered by the recalling firm's CAPA: 1) Burrs were found to be present on Left Housing, Right Housing and Barrel components, 2) Loose plastic particulate was found within sealed primary packages, 3) Loose hair was discovered within four sealed primary packages and 4) Assembly error, wherein Left and Right Housing components were acceptably press-fit together, was discovered within one sealed primary package. One of the Left Housing press-fit pins was not pressed into the associated hole of the Right Housing, resulting in a small gap in the housing assembly.